Pular para o conteúdo principal
GE Healthcare · CTLightSpeed23001608730636476964096230016001433713312230005088

LightSpeed 2.X (Plus, QX/i) - Octane

Symptom

Images fail to display and recon shutdown. Images fail to display and recon shutdown

Error message

Error: 230016087 Call to DMA library failed See LD for more details Wed Jun 21 08:48:13 2006 Error: 230016087 Host: ice-post Process: tsdc0 File: dma_chain.c Line: 4278 Tower 0, Dsp_id 3 Function Current State : 0x0 Time Stamp : 30636 sec : 476964096 nsec PATH :prep_apps.c:630/prep_view_loop.c:229 INFO : timed out waiting for link dma handle to become inactive 0x00000000 0x00000000 0x00000388 Call to DMA library failed

Error: 230016001 prep get cal failed See long descripton for an example error message. Wed Dec 2 08:17:53 2009 Error: 230016001 Host: ice-post Process: img0 File: prep main.c Line: 334 Tower 1, Dsp id 2 Function Image Generation : Preprocessing Current State : 0x0 Time Stamp : 1035 sec : 433713312 nsec PATH : INFO : sharc cal cs=0x7fffffff:ppc cal cs=0x7fffffff; sharc offset cs=0x7fffffff:ppc offset cs=0x7fffffff prep get cal failed

Error: 230005088 Msg Queue recieve failed for dsp = 2 on tower = 1 with status = 10. See long description for an error example. Wed Dec 2 08:17:38 2009 Error: 230005088 Host: ice-post Process: img0 File: dsp class.cxx Line: 200 Msg Queue recieve failed for dsp = 2 on tower = 1 with status = 10.

Applies to

Models
LightSpeed
Product
LightSpeed 2.X (Plus, QX/i) - Octane Always S-DAS 2.4.2_H2.4M5 LightSpeed 2.X (Plus, QX/i) - Octane (Pegasus or Rhapsode IG) - Always S-DAS Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 2.4.2_H2.4M5 FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 FEATURES/OPTIONS STD/OPTION Cat # Digital Laser Camera I/F B7700PD Smart Helical (s/w only) Standard B7530TK Fast Recon (s/w only) Option B7700JF BMD (requires B7710MP or MN) Option B7710MK Smart Prep Option B7530SP Smart Score ? ? 3D Analysis (Requires D3D) Option B7540TD Navigator (s/w only) Option B7540N DentaScan (s/w only) Option B7540SD Direct 3D (Requires B7710SA) Option B7810SS HIS/RIS Option (h/w and s/w) Option B7500PN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2216467 Pegasus Image Generator used on LS 1.x-4.x Octane, GOC1 and GOC2. Not GRE See long description for details Part #: 2216467 Part Type: GP REPAIRABLES Part Weight: 4.41 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: FAST IMAGE GENERATOR FRU: Yes Status: Orderable Comments: FLS - PEGASUS IMAGE GENERATOR PWA CIRCUIT BOARD. USED ON HELIOS 2. MS26561 DSM 2/24/01 PCN 220318 ERRONEOUSLY CHANGEDBOARD VERSION FROM E TO G (REVISION G IS WRONG, NOTHING SHOULD BE MANUFACTURED UNDER THAT REVISION, PER PCN 220318). ADDED BY MS20544, CSV ON 1/21/02. 7/19/02- C2R COMPLETED. MS12464 4/21/03 - FEILD RESTOCK FLAG RESET TO Y. MS12575 LIST PRICE $26371.00 GELS - SECOND OPINION PROCESS TO ASK OLC IN BUC BEFORE TO SHIP INFO FROM P DE LA FUENTE /14/06/TS GA0312 EN ATTENTE DE REGUL STOCK CONTACTER GILLES ANCELOT SH CSO 0300690 ITALIE I 24810431002,PCB PEGASUS IMAGE GENERATOR PWA See example below for GOC1/GOC2 ICE box with pegasus IG board.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required