Pular para o conteúdo principal
GE Healthcare · CTLightSpeed18370544533182119

LightSpeed 2.X (Plus, QX/i) - Octane

Symptom

System wont boot up. Stops at OC initializing. System wont boot up. Stops at OC initializing

Error message

Error: 183705 Console Push Button Watchdog Timeout. See LD for more details Sat Nov 25 04:49:14 2006 Error: 183705 Host: STC Process: Console Pushbuttons File: cp_wdog.c 1.23 Line: 215 Function: No System Function Reported Scan: 44533/2/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 04:49:11:650 Log Series: 353 Console Push Button Watchdog Timeout. HW Error: Possible hardware problem.

Error: 182119 LAN Watchdog TIMEOUT for the SBC. See Long Description for more details Sun Sep 17 10:24:42 2006 Error: 182119 Host: STC Process: Master Exc Handler File: exm_lan_wdog.c Line: 556 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 10:24:20:560 Log Series: 37 LAN Watchdog TIMEOUT for the SBC. HW/SW Error: Did not receive LAN watchdog message.

Applies to

Models
LightSpeed
Product
LightSpeed 2.X (Plus, QX/i) - Octane Always S-DAS 2.4.2_H2.4M5 LightSpeed 2.X (Plus, QX/i) - Octane (Pegasus or Rhapsode IG) - Always S-DAS Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 2.4.2_H2.4M5 FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 FEATURES/OPTIONS STD/OPTION Cat # Digital Laser Camera I/F B7700PD Smart Helical (s/w only) Standard B7530TK Fast Recon (s/w only) Option B7700JF BMD (requires B7710MP or MN) Option B7710MK Smart Prep Option B7530SP Smart Score ? ? 3D Analysis (Requires D3D) Option B7540TD Navigator (s/w only) Option B7540N DentaScan (s/w only) Option B7540SD Direct 3D (Requires B7710SA) Option B7810SS HIS/RIS Option (h/w and s/w) Option B7500PN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2260421 replaces 2260421-2 VME console power supply. Used on CTi - 4.x H3 GOC1 and GOC2 Operators Console. See Long Description for Details. 2260421 replaces 2260421-2 Part #: 2260421 Part Type: GP REPAIRABLES Part Weight: 2.18 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: VME CONSOLE POWER SUPPLY ASSEMBLY FRU: Yes Status: Orderable Part number 2260421-2 is a comsumable assembly at EOL with 45 per year comsumsion. Part number 2260421 is the same base assembly with an added bracket and is a repairable. Do to the EOL issue and low comsumption of 45 per year we will eliminate p/n 2260421-2 and replace it with p/n 2260421 to have one repairable part. This will eliminate the need to stock two parts and create a new repair process.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues