Pular para o conteúdo principal
GE Healthcare · MRSigna2218078

Images Not Displayed in the Viewer or the Browser due to disk drive.

Symptom

Images Not Displayed In The Viewer Or The Browser

Error message

2218078 Cannot find sdc path for image #1 Wed Dec 2 08:39:33 2009 Error: 2218078Host: tcmri1 Process: TIRFile: ImagePath.c Line: 79Exception Class: UnknownFunction: Host: IFCCCannot find sdc path for image #1

I/O error in filesystem ("sda2") meta-data dev sda2 block 0x4ff972.

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
5183547-15 73GB SAS 15K RPM Disk Drive 5183547-15 73GB SAS 15K RPM Disk Drive
Area
MR