Excessive cutoff on one edge, typically when the C-arm is in the horizontal plane.
Symptom
Excessive cutoff on one edge, typically when the C-arm is in the horizontal plane. See Long Description The D5K has excessive cutoff on one edge, typically when the C-arm is in the horizontal plane. The X-ray Tube (XRT) to Detector C-arm alignment in the vertical plane seems proper, but the collimator light field, crosshairs, and red laser do not line up with the detector cover template center indicators. The X-ray Tube beam center tends to drift/shift when you change from the 100cm SID position to the 180 cm SID position (see Step# 7).
Applies to
- Models
- Definium
- Product
- Definium 5000
- Area
- X-ray R&F