GE Healthcare · MRSigna22630785063322630752254415
Will Not Scan With RF Errors due to SRFD2.
Symptom
Will Not Scan With RF Errors Will not scan with RF errors
Error message
2263078 A software transmit overflow occurred on a UART link Mon Dec 11 12:47:09 2006 Error: 2263078 Host: MR1 Process: NSP File: SCP:../uartLib.c Line: 343 Exception Class: Unknown Severity: Unknown Function: Peripherals: Communications PSD: Spin Echo Coil: BODY Scan: 50633/1 UART Communication Error A software transmit overflow occurred on a UART link. Source: RF NB Amp Interface UART: Debug UART 2263075 UART Communication Error A software receive overflow occurred on a UART link 2263075 NSP SCP:../uartLib.c 440 Exception Class: Unknown Severity: Unknown Function: Peripherals: Communications PSD: Unknown Coil: Unknown Scan: -1/-1 UART Communication Error A software receive overflow occurred on a UART link. Source: RF NB Amp Interface UART: Debug UART 2254415 Attention: RF Amplifier in service, not product mode 2254415 NSP SCP:../rf_debug.c 443 Exception Class: Unknown Severity: Unknown Function: RF: RF Transmit PSD: Unknown Coil: Unknown Scan: -1/-1 Attention: RF Amplifier in service, not product mode. The RF Amplifiers are being controlled via the serial terminal. Reset the RF Amplifier Interface or Reset TPS to return control to the system.
Applies to
- Models
- Signa
- Product
- Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
- Software
- 15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
- Component
- SecondOpinion SRFD2 1.5T Solid State RF Amplifier 2351573 See comments in Resolution Concept for second opinion requirements before ordering this part. SecondOpinion SRFD2 1.5T Solid State RF Amplifier 2351573
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.