Symptom
Intermittently table is not available for a short time (about25 sec). Cradle symbol of gantry display is blinking during that time. Operator cannot do any move table.
Error message
Error: 260116313 Table Gantry Processor - Host (OC) LAN Communication Watchdog Timed out. see LD Tue May 16 22:03:29 2006 Error: 260116313 Host: tgp Process: tgp_subsys File: subsys_mgr.c Line: 349 Running: System Reset Tue May 16 21:37:53 2006 Host : Table Gantry Processor Ermes # : 260116313 Exception Class : Safe State Severity : Secondary File : subsys_mgr.c Line# : 349 Function : SYSTEM COMMUNICATIONS : Control Network Scan Type : None/Unknown Scan: 0/0/0 Table Gantry Processor - Host (OC) LAN Communication Watchdog Timed out. Error: 260116316 Communications are DOWN between subsystems See LD for full error Mon Jun 13 14:34:09 2005 Error: 260116316 Host: tgp Process: tgp_subsys File: lan_watchdog.c Line: 904 Mon Jun 13 14:34:03 2005 Host : Table Gantry Processor Ermes # : 260116316 Exception Class : Safe State Severity : Primary File : lan_watchdog.c Line# : 904 Function : No System Function Reported Scan Type : None/Unknown Scan: 7754/5/0 Communications are DOWN between subsystems Or Sun Dec 17 08:26:45 2006 Error: 260116316 Host: tgp Process: tgp_subsys File: lan_watchdog.c Line: 996 Sun Dec 17 08:26:39 2006 Host : Table Gantry Processor Ermes # : 260116316 Exception Class
Applies to
- Models
- LightSpeed
- Product
- LightSpeed RT16 & RT16 Xtra and Vision WB_H_G16_G_GTL = RT16, Xtra WB_H_G16_G_GTL650 = Vision WB_H_G16_G_GTM = RT16, Xtra or Vision LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry Links to the Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5139590 Control assembly left side gantry. ( replaces part# 2256729 and 2330695 ) – used on H2 & HP gantry’s (LightSpeed 2.x, 3.x, 4.x, 5.x, 7.x ) FEMC info in Long Description. 5139590 Control assembly left side gantry. ( replaces part# 2256729 and 2330695 ) – used on H2 gantry’s (LightSpeed 2.x, 3.x, 4.x, 5.x, 7.x ) Part #: 5139590 Part Type: GP CONSUMABLES Part Weight: 4.41 LBR Repairable: No Returnable: No Hazardous Material: N Description: H POWER OPERATOR CONTROLS LEFT SIDE FRU: Yes Status: Orderable If part number 2330695 is ordered it will automaticaly cross to part 5139590 Old part info 2256729 Control assembly left gantry Part #: 2256729 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: CONTROL ASM LEFT H2 GANTRY FRU: Yes Status: Orderable 5139591 control assembly right side gantry ( replaces part# 2256728 and 2330703 ) ) – used on LightSpeed 2.x, 3.x, 4.x, 5.x, 7.x (HD uses a different control panel) See Long Description. 5139591 control assembly right side gantry ( replaces part# 2256728 and 2330703 ) ) – used on H2 gantry,s (LightSpeed 2.x, 3.x, 4.x, 5.x, 7.x ) Part #: 5139591 Part Type: GP CONSUMABLES Part Weight: 2.20 LBR Repairable: No Returnable: No Hazardous Material: N Description: H POWER OPERATOR CONTROLS RIGHT SIDE FRU: Yes Status: Orderable OLD part info 2256728 control assembly right gantry Part #: 2256728 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: CONTROL ASM RIGHT H2 GANTRY FRU: Yes Status: Orderable Comments: If no stock is available, order 5139591 per ECR 2025944; 6/20/2006 MS11892 FLS - PLEASE CONVERT SPH ORDER TO OVERNIGHT SHIPMENT IF PART AVAILABLE IN MEMPHIS AND IF FEDEX SERVICE (INCLUDING 8:00 AM SERVICE) MEETS FE NEED TIME CRITERIA Part #: 2330703 Part Type: GP CONSUMABLES Part Weight: 2.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: H POWER OPERATOR CONTROLS RIGHT SIDE FRU: Yes Status: Orderable Parts Offered as: Current: 2330703 Harvested: Not Available Exchange: Not Available Comments: If no stock is available, order 5139591 per ECR 2025944; 6/20/2006 MS11892. 212012201 02/09/2007 19:34 GMT
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.