Pular para o conteúdo principal
GE Healthcare · CTLightSpeed26000745750084

LightSpeed 4.X - GRE

Symptom

Collimator Cal fails in fastcals. Collimator Cal fails in fastcals

Error message

Error: 260007457 Beam Tracking Control Loop Error Exceeded The ABORT Tolerance See Long Description for complete error Wed Apr 27 18:54:40 2005 Error: 260007457 6 DAS DCB Firmware zaxis.c 564 Function: DATA ACQUISITION : Scan Data Acquisition Scan: 50084/6/0 Type: None/Unknown Exception Level: Pri/Soft Time: 18:54:40:490 Log Series: 936 Beam Tracking Control Loop Error Exceeded The ABORT Tolerance. Tolerance : 400 umd Control Error : 448 umd Beam Side : B Side

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues