Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT18200450238431109430857

LightSpeed 4.X - GRE

Symptom

LAN watchdog Errors

Error message

Error: 182004 LAN Watchdog TIMEOUT See LD for the full error code Mon Feb 14 10:28:30 2005 Error: 182004 Suite: SJ01 Host: ETC Process: Local Exc Handler File: exl_lan_in_mail.c Method: No Method Line: 175 Function: No System Function Reported Scan: 50238/1/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 10:28:30:550 Log Series: 31 LAN Watchdog TIMEOUT. HW/SW Error: Did not receive LAN Watchdog message from STC. (Current Time:431109 Last Wdog LAN Probe Time:430857 (in Ticks))

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues