GE Healthcare · CTLightSpeedVCT260007422260007318
LightSpeed 4.X - GRE
Symptom
Can't scan with Pink Warning: Software has detected an invalid DAS configuration.
Error message
Error: 260007422 Data Parity Error detected in the 5B Data Stream on Converter Board 2 or the DCB Board. See long description Mon Feb 1 06:34:22 2010 Error: 260007422 Host: DAS Process: DCB Firmware File: parity_16.c Line: 247 Function: DATA ACQUISITION : Scan Data Acquisition Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 06:34:23:235 Log Series: 31 SCAN ABORT -- DAS Has Detected An Error.^B Data Parity Error detected in the 5B Data Stream on Converter Board 2 or the DCB Board. (CSI:1 RSM:0x1303). Error: 260007318 DCB Power Supply self-test failed. See LD for more details Fri Mar 3 09:02:35 2006 Error: 260007318 Host: DAS Process: DCB Firmware File: hdwtest.c Line: 135 Function: DATA ACQUISITION : Scan Data Acquisition Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 09:02:33:520 Log Series: 34 A Failure Has Been Detected By A DCB Selftest.^B DCB Power Supply self-test failed.
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required