Pular para o conteúdo principal
GE Healthcare · CTLightSpeed244587244584

LightSpeed 4.X - GRE

Symptom

Scan Hardware is not available message appears on the monitor WHEN clicking Tube Warm-up. Scan Hardware is not available message appears on the monitor WHEN clicking Tube Warm-up

Error message

Error: 244587 following subsystems failed to report initialization status in 50 seconds OBC SS(Coll/Das). Error: 244587 Host: qxi16 Process: fwmgr File: HwSetUp.c Line: 2095 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 The following subsystems failed to report initialization status in 50 seconds: OBC SS(Coll/Das).

Error: 244584 Collimator/DAS reset failed See Long Description for complete error Error: 244584 Host: qx-ct Process: fwmgr File: HwManager.c Line: 1536 Collimator/DAS reset failed.

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5191271 MDAS 16 Fuse Holder Upgrade Kit. Supercedes part number 46-170042P28. See Long Description for more information. Part #: 5191271 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: MDAS 16 Fuse Holder Upgrade Kit FRU: Yes Status: Orderable Comments: Per ECR 2036788, 212008898, 11/30/2006. MDAS Fuse Holder Replacement Kit Contents: Part #: 46-170042P28 Part Type: GP CONSUMABLES Part Weight: 2.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: 30A, 250V PANEL-MOUNT FUSEHOLDER. FRU: Yes Status: Orderable Comments: FLS - FUSE HOLDER PART OF FMI KIT ** FE Assist ** IF FUSE HOLDER P/N 46-170042P28 FAILED, PLEASE ORDER MDAS FUSE FIX ON FAILURE KIT 5191271 (12/12/06 212001261)
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.