Pular para o conteúdo principal
GE Healthcare · CTDiscoveryVCT26013279350089

Scan hardware stopped scan with DAS Board errors.

Symptom

Scan hardware stopped scan with DAS Board errors. Scan hardware stopped scan DAS Board errors Scanner Hardware Stopped Scan message appears in the operator console area during Axial Scan data acquisition with Error: 260001023 The DCB has notified the system of an error condition which prevents scanning and error: 260007402-Converter Board(s) Asserted a Board FAULT cards 76,78,80,82,84,86,88,90,92,94,96 and Error #260007774 list of converter cards from previous error message Problem was intermittent Then after time the errors would occur more often and eventually the problem was consistent [insite@erct ~]$ /usr/g/insite/bin/viewlog Sun Aug 29 Error: 260001023 Host: OBC Process: HOP File: hop erlog.c Line: 367 Function: No System Function Reported Scan: 50000/1/2 Type: Axial Exception Level: Pri/Most Severe Time: 08:37:10:145 Log Series: 120 The DCB has notified the system of an error condition which prevents scanning. See log for further information:^C Error message code: 260002358 Notification level: 5 Sequence number: 11

Error message

Error: 260132793 Digital Modules Have Not Been Detected By The Data Acquisition System See Long Description for complete error. Mon Feb 6 09:42:29 2006 Error: 260132793 Host: orp Process: orp_subsys File: IfbManager.cxx Line: 318 Mon Feb 6 09:42:28 2006 Host : Das Ermes # : 260132793 Exception Class : Abort Severity : Primary File : IfbManager.cxx Line# : 318 Function : No System Function Reported Scan Type : Static Scan: 50089/1/1 The Following Digital Modules Have Not Been Detected By The Data Acquisition System. See Previous Messages From Last Reset For Detailed Fault Information. Digital Modules: 37-40 (or some equivalent number)

Applies to

Models
Discovery, VCT
Product
Discovery CT750 HD V40_PS_HD64_G_GTL V40_PS_HD64_G_GTM Discovery HD CT750 Method of visual identification HD gantry: Gray top and Gray Side covers with HD Markings Console Type: GOC6 has black SCIM and tower face Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of softwares the product can run (Consult attached FCT Software Cross Reference): 08MW39.4 - M3 (see attachement for M3 available options) 09MW29.14 - M4 (HD Options = M3 plus Gemstone Spectral Imaging, Gemstone Spectral Review and volume Helical Shuttle ) Example: Of difference with options software [insite@ct01 ~]$ swokinstall -p Product Software Option Keys: Gemstone Spectral Imaging Permanent HelicalShuttle Permanent Xtream Recon Permanent FAQ's Table weight limit: 500 lbs Tube Type - D3702T Green D3701T Current production - M4 Refer to product data sheets or Technical reference Manual, English 5317222-1EN Revision : 2 or later Refer to Application Tips and Work-Arounds Direction 5311815-1EN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required