GE Healthcare · CTLightSpeed18415150004245190
LightSpeed 2.X (Plus, QX/i) - Octane
Symptom
Scan hardware stopped or reset during Fastcals. Scan hardware stopped or reset during Fastcals
Error message
Error: 184151 Scan Control timed out waiting for the Ready For Views message from the SRU See the long description for the full error code Sat Apr 17 20:29:44 2004 Error: 184151 Host: STC Process: Scan Control File: scTrigSeqActions.c 1.65 Line: 1365 Function: DATA ACQUISITION : Scan Data Acquisition Scan: 50004/3/1 Type: Axial Exception Level: Pri/Most Severe Time: 20:29:43:355 Log Series: 676 Scan Control timed out waiting for the Ready For Views message from the SRU. This message is sent from the SRU once offset collection has completed. Possible causes: trigger failure, DIP failed to receive data, message lost, software error Error: 245190 System aborted Axial scanning. No views, either offset or data, were collected/received. Data views expected: 992 See the Long Description for the full error code. Thu Jan 13 06:58:23 2005 Error: 245190 Host: ct02 Process: scanmgr File: scan_complete.cxx Line: 303 Function: System Control : Scan Control Scan Type: Axial Scan: 0/0/0 Scan Seq Id: 1000 System aborted Axial scanning. No views, either offset or data, were collected/received. Data views expected: 992
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 2.X (Plus, QX/i) - Octane Always S-DAS 2.4.2_H2.4M5 LightSpeed 2.X (Plus, QX/i) - Octane (Pegasus or Rhapsode IG) - Always S-DAS Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 2.4.2_H2.4M5 FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 FEATURES/OPTIONS STD/OPTION Cat # Digital Laser Camera I/F B7700PD Smart Helical (s/w only) Standard B7530TK Fast Recon (s/w only) Option B7700JF BMD (requires B7710MP or MN) Option B7710MK Smart Prep Option B7530SP Smart Score ? ? 3D Analysis (Requires D3D) Option B7540TD Navigator (s/w only) Option B7540N DentaScan (s/w only) Option B7540SD Direct 3D (Requires B7710SA) Option B7810SS HIS/RIS Option (h/w and s/w) Option B7500PN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2169940-45 IP-30 used on Octane I Computer See LD for more info 2169940-45 IP-30 used on Octane I Computer Part #: 2169940-45 Part Type: GP REPAIRABLES Part Weight: 1.81 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: PCA REV F HEART IP30 SYSTEM BOARD FRU: Yes Status: Orderable Comments: FLS - PCA REV F HEART IP30 SYSTEM BDU/O HELIOS & CTI FOR CTI SYSTEMS MUST HAVE 6.1 SOFTWARE SCHEDULED FOR RELEASE FW 29 2000 PER PCN 199811 MS08992 6/27/00 MS5500 6/21/01 SGI NUMBER 030-1467-001. CHECKLABEL ON BOARD TO DETERMINE YOUR BOARD TYPE. IF BOARD BEING REPLACED IS LABELED 030-0887-005 ORDER THE T3202KT !!!! IF SYSTEM CAN BE BROUGHT UP TOCOMMAND MONITOR, THE SYSTEM COMMAND WILL LIST IP30 BOARD TYPE. IF SYSTEM CAN BE BROUGHT UP TOIRIX, THE HINV -M COMMAND WILL LIST IP30 BOARD TYPE. PLEASE ORDER THE CORRECT TYPE OF BOARD TO PREVENT COMPATIBILITY ISSUES AND INCREASED COSTS. GELS - "PCA REV F HEART IP30 SYSTEM BD U/O HELIOS & CTI PCN 199575 BACKWARD COMPATIBLE WITH 2169940-13 (ALL GONE') SEE D91SOURCE: SGI-BROOKFLIED, WI (SILICON GRAPHICS, INC.) MS08992 8/25/99",IP 30 BD Note: Octane 1 has two ESI Graphics boards ct01 10# hinv -mv Location: /hw/node PM20300MHZ Board: barcode LLJ521 part 030-1356-001 rev N Location: /hw/node/xtalk/15 IP30 Board: barcode LKF767 part 030-1467-001 rev D Location: /hw/node/xtalk/15/pci/2 FP1 Board: barcode MCC875 part 030-0891-003 rev H PWR.SPPLY.ER Board: barcode AAE1060336 part 060-0035-002 rev A Location: /hw/node/xtalk/13 XTALKPCI Board: barcode LKV989 part 030-0952-005 rev E Location: /hw/node/xtalk/13/pci/3 PCI_ENET Board: barcode KZM587 part 030-1155-002 rev D Location: /hw/node/xtalk/12 MOT10 Board: barcode LDX145 part 030-1241-002 rev K Location: /hw/node/xtalk/11 MOT10 Board: barcode LKJ740 part 030-1241-002 rev K 2 300 MHZ IP30 Processors Heart ASIC: Revision F CPU: MIPS R12000 Processor Chip Revision: 2.4 FPU: MIPS R12010 Floating Point Chip Revision: 0.0 Main memory size: 1536 Mbytes Xbow ASIC: Revision 1.4 Instruction cache size: 32 Kbytes Data cache size: 32 Kbytes Secondary unified instruction/data cache size: 2 Mbytes Integral SCSI controller 0: Version QL1040B (rev. 2), single ended Disk drive: unit 1 on SCSI controller 0 (unit 1) Disk drive: unit 2 on SCSI controller 0 (unit 2) Integral SCSI controller 1: Version QL1040B (rev. 2), single ended Optical disk: unit 3 on SCSI controller 1 CDROM: unit 6 on SCSI controller 1 Integral SCSI controller 2: Version QL1040B (rev. 2), single ended Disk drive: unit 1 on SCSI controller 2 (unit 1) IOC3 serial port: tty1 IOC3 serial port: tty2 IOC3 parallel port: plp1 Graphics board: ESI with texture option Graphics board: ESI Integral Fast Ethernet: ef0, version 1, pci 2 Fast Ethernet: ef1, version 1, pci 3 Iris Audio Processor: version RAD revision 12.0, number 1 PCI Adapter ID (vendor 4265, device 3) pci slot 2 PCI Adapter ID (vendor 4215, device 4128) pci slot 0 PCI Adapter ID (vendor 4215, device 4128) pci slot 1 PCI Adapter ID (vendor 4265, device 5) pci slot 3 PCI Adapter ID (vendor 4215, device 4128) pci slot 1 PCI Adapter ID (vendor 4431, device 40) pci slot 2 PCI Adapter ID (vendor 4265, device 3) pci slot 3 PCI card, bus 0, slot 2, Vendor 0x114f, Device 0x4
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required