GE Healthcare · CT
Lightspeed 1.X - Octane
Symptom
System will not boot to or stay in applications. System will not boot or stay in applications
Error message
ALERT: PIO Error on XIO Bus /hw/node/xtalk/0/xbow [xbow#0] port 10 This error was seen in /var/adm/SYSLOG. See long description for more information. This error was seen in /var/adm/SYSLOG as part of a longer string of error messages there. Below is the complete stream of errors that were seen in the log at the time: May 16 10:38:30 2E:gect savecore: reboot after panic: PANIC: KERNEL FAULT May 16 10:38:30 2E:gect savecore: pb 19: 1 ALERT: PIO Error on XIO Bus /hw/node/xtalk/0/xbow [xbow#0] port 10 May 16 10:38:30 2E:gect savecore: pb 20: 6 ^IAccess attempted to offset 0x4 May 16 10:38:30 2E:gect savecore: pb 21: 6 ^ILink failed while transferring data May 16 10:38:30 2E:gect savecore: pb 22: 4 WARNING: Xbow at /hw/node/xtalk/10 encountered Fatal error May 16 10:38:30 2E:gect savecore: pb 23: 6 xtalk PIO Read error in kernel mode May 16 10:38:30 2E:gect savecore: pb 39: EXC code:28, `Data Bus Error '
Applies to
- Product
- Lightspeed 1.X - Octane 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete Lightspeed 1.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config showprods|grep Apps note: if check_config does not work then the software is not up to date with all FMI's. Check FMI history and verify that all patches have been loaded since last software reload. Ref FMI 25269 List of Supported Console Types Octane with Pegasus Recon hardware. may have an 02 SBC - Run df -k to see if the sbc is partitioned. List of softwares the product can run 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2225004-9 73GB scan disk (replaces 2225004-8 and 18 GB p/n 2225004-4 and 36 GB scan disk 2225004-5) See Long Description for Details. Part #: 2225004-9 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: 73GB 15000 RPM Scan Data Drive FRU: Yes Status: Orderable Comments: Per ECR 2046173, 212008898, 8/8/07. ** FE ASSIST ** Refer to Service Note SN07110612a For Jumper information (12/3/07 212001261) Part #: 2225004-8 Part Type: GP CONSUMABLES Part Weight: 1.61 LBR Repairable: No Returnable: No Hazardous Material: N Description: HARD DRIVE 73GB 15KRPM SCAN DATA FRU: Yes Status: Orderable Comments: Added per ECR 2031409, 7/14/06, ms11892. ** FE ASSIST ** If no stock of the 2225004-8 order 2225004-9 as the replacement. Refer to Service Note SN07110612a (212001261 12/3/07)
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required