GE Healthcare · MRSigna222466622523012252293
Multicoil TR driver fault on Neo 3.0T HD 8 Channel Shoulder Coil due to a design problem.
Symptom
Unable to Scan with Multi Coil TR Driver Faults
Error message
2224666 Auto Prescan failed, Driver Module Error Occurred. Wed Apr 21 20:00:24 2004 Error: 2224666 Host: lx-mr Process: psc File: PscMainControl.cc Line: 1334 Auto Prescan failed, Driver Module error occurred. 2252301 Multicoil Bias Fault Detected - Open Circuit. Try re-seating all coil connectors. MCDx Channel x of y, counting from 1 Value x mA Limit y mA See Long Description for Extended Error 2252301 MRI_GB_OW0 NSP SCP:DCB 816 Multicoil Bias Fault Detected - Open Circuit. Try re-seating all coil connectors. MCD1 (or MCD2) Channel 1 of 16, counting from 1 Value: 2 mA Limit: 298 mA Mask of failed channels: 0x1 Ch 1-8 : Legacy MC Conn Ch 17-24 : Conn A Ch 25-32 : Conn B-> Extended Error: This fault was detected by the Driver Module. An open circuit fault occurs when the actual value is less than the pre-defined limit. Check system cables, connectors, coils and configurations. Re-seat all coil connectors. The mask of all failed channels indicates additional channels with the same fault - convert the hex number to a bitmap to decipher the mask, with channel 1 indicated by the least significant bit and channel 16 indicated by the most significant bit. For example, a mask of 0x3c conv
Applies to
- Models
- Signa
- Product
- 3.0T Signa HDxT For 3.0T Signa HDxT systems with 15.0 software only. For 3.0T Signa HDxT systems with 15.0 software only.
- Software
- 15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
- Component
- 2414331 3.0T Signa HDx Shoulder Coil
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.