Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT230007008

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

Pop up during fastcal indicates potential image quality issue. See Long Description for error stream example. Pop up error posted to operator during fastcal indicated potential image quality issue. Errors 230007000 230007002 230007014 posted in gesyslog. Exclusion test, open channel failed Z-axis exclusion and proximity criteria. Fri Oct 28 09:32:03 2005 Error: 230007000 Host: presbyplano Process: badchanmgr File: BadChannelMap.cxx Line: 616 Exclusion Test - Row 64:32B, Channel 695, Module 44 Open Channel* FAILED Z- AXIS EXCLUSION CRITERIA *. Fri Oct 28 09:32:03 2005 Error: 230007002 Host: presbyplano Process: badchanmgr File: BadChannelMap.cxx Line: 638 Exclusion Test - Row 64:32B, Channel 695, Module 44 Open Channel* FAILED PR OXIMITY CRITERIA *. Fri Oct 28 09:32:03 2005 Error: 230007014 Host: presbyplano Process: badchanmgr File: badchanmgr.cxx Line: 1194 Uncorrectable detector channels have been found for scan 50029/2/2

Error message

Error: 230007008 Exclusion Test - Row 33:1B, Channel 912, Module 57 Short to ground* FAILED PROXIMITY CRITERIA * Sat Sep 17 08:08:03 2005 Error: 230007008 Host: amhvct Process: badchanmgr File: BadChannelMap.cxx Line: 652 Exclusion Test - Row 33:1B, Channel 912, Module 57 Short to ground* FAILED PROXIMITY CRITERIA *.

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
5134959 FRDM Module L-FINE 32 replaces 5123019 Used on VCT 32 Detector See Long Description for Details. 5134959 FRDM Module L-FINE 32 replaces 5123019-Used on VCT 32 Detector Part #: 5134959 Part Type: GP CONSUMABLES Part Weight: -98999.01 LBR Repairable: No Returnable: No Hazardous Material: N Description: L-FINE 32 Detector Module Kit FRU: Yes Status: Orderable Part #: 5123019 Part Type: MST DEFAULT LOAD Part Weight: 0.00 LBR Repairable: Not Available Returnable: Not Available Hazardous Material: N Description: L-FINE 32 DIGITAL MODULE ASSY MP3 FRU: Yes Status: Non Orderable
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues