Pular para o conteúdo principal
GE Healthcare · CTLightSpeed

LightSpeed 2.X (Plus, QX/i) - Octane

Symptom

Images are showing up on the monitor with autoview but not in Image Works. Archiving images and networking of the images are failing.

Error message

ALERT: I/O error in filesystem ("/usr/g/sdc image pool2") meta-data dev 0xed block 0xa00040 ("xfs trans read buf") seen in /var/adm/SYSLOG see LD ALERT: I/O error in filesystem ("/usr/g/sdc image pool2") meta-data dev 0xed block 0xa00040 ("xfs trans read buf") seen in /var/adm/SYSLOG The I/O error could report to a number of different file systems / disk drives. i.e. ("sda8")

Error: 9065 Internal error, fork not allowed NO FREE PID See LD for complete error Tue Sep 12 17:50:38 2006 Error: 9065 Host: sjnect2 Process: dbwserver File: proc.c Line: 187 Internal error, fork not allowed NO FREE PID

dksc0d2s0: 6 SCSI driver error: device does not respond to selection seen in /var/adm/SYSLOG

Applies to

Models
LightSpeed
Product
LightSpeed 2.X (Plus, QX/i) - Octane Always S-DAS 2.4.2_H2.4M5 LightSpeed 2.X (Plus, QX/i) - Octane (Pegasus or Rhapsode IG) - Always S-DAS Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 2.4.2_H2.4M5 FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 FEATURES/OPTIONS STD/OPTION Cat # Digital Laser Camera I/F B7700PD Smart Helical (s/w only) Standard B7530TK Fast Recon (s/w only) Option B7700JF BMD (requires B7710MP or MN) Option B7710MK Smart Prep Option B7530SP Smart Score ? ? 3D Analysis (Requires D3D) Option B7540TD Navigator (s/w only) Option B7540N DentaScan (s/w only) Option B7540SD Direct 3D (Requires B7710SA) Option B7810SS HIS/RIS Option (h/w and s/w) Option B7500PN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2169940-72UU (replaces 2169940-72) : Octane Computer 9 GB System Disk See Long Description for details Note: There are compatability problems with some consoles and disk drive firmware. Please see service note SN07112612a for details. Part #: 2169940-72UU Part Type: GP USED Part Weight: 2.43 LBR Repairable: No Returnable: No Hazardous Material: N Description: 9GB 10K RPM DISK WITH SLED FRU: Not Available Status: Orderable Comments: NOTE: MAY BE INACCURATE FOR H2 AND H3 CONSOLES. ** FE ASSIST ** 2169940-12 is no longer available can use 2169940-72 9GB or a 2169940-78 18GB Hard drives for a CTi or lightspeed QXI Octane systems as Image or OS disk. Will need to use updated LFC per 2282382-100 rev 2 or 2284291-100 rev 8. The 2169940-78 has 5000 images files where normal is 20000 if used as a CTi Image Disk. Refer to Service Note 07112612b. (212001261 8/11/09) ***OBSOLETE PART Case # 9746502*** This part has been identified as technically obsolete and no short term solution has been identified. FE must identify and order a higher or lower level spare as required. Contact your support engineering, On-line Engineering or Parts ID team should you require support. 212012173 7-24-236 Part #: 2169940-72 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 1.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: 9GB Octane 10K RPM DISK WITH SLED FRU: Yes Status: Orderable Parts Offered as: Current: 2169940-72 Used: Not Available Exchange: Not Available Comments: OCTANE 2 9GB DISK DRIVE USED ON HELIOS 2 AND 3 AND LX 9/25/03 CHECK WITH FIELD TO SEE IF THEY CAN USE THE 18GB PART# 2169940-78 IF NONE OF THE 9GB ARE AVAILIABLE
Area
CT

Full cause & step-by-step resolution