GE Healthcare · MRSigna2261176

Unable to Download or Reset the MGD Chassis with Error 2261176 due to 16 Channel Switch.

Symptom

Unable to Download or Reset the TPS MGD or ISE Unable to download TPS or ISE or MGD chassis

Error message

2261176 SRI to 16 Channel Switch Communication Error: An error occurred reading the serial ID on the 16 Channel Switch. 2261176 tempmr NSP SCP:SriAutoConfig 796 SRI to 16 Channel Switch Communication Error: An error occurred reading the serial ID on the 16 Channel Switch. Please check the connection between the SRI and switch and make sure the switch has power. 2261176 tempmr NSP SCP:SriAutoConfig 796 SRI to 16 Channel Switch Communication Error:An error occurred reading the serial ID on the 16 Channel Switch. Please check the connection between the SRI and switch and make sure the switch has power.

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
SecondOpinion 1.5T Excite HD 16 Channel Switch 5109645 See comments in resolution concept for second opinion requirements before ordering this part
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.