GE Healthcare · MRSigna2225385
MOD Drive will not attach Due to no power to expansion chasis.
Symptom
MOD Drive will not attach for archive. The system is a 5X Horizon system that was upgraded to use the Maxoptix MOD drive and media. An FE tried to order the Maxoptix T5-2600 drive thru Cares, but due to end of life issues, the drive changed up to a Sony MOD drive. After installing the new Sony drive, the system would recognize the drive as a Sony but could not attach or archive using the drive.
Error message
2225385 Workspace Interface Module Initialize failed. See Long Description Mon May 23 08:17:18 2005 Error: 2225385 Host: lx-mr Process: NSP File: HPC_PORT_ID:spi_cpb_initialize_fail Line: 4 Workspace Interface Module Initialize failed. If recovery fails, check Workspace Interface Module connections and run Workspace Interface Module diagnostics.
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.