GE Healthcare · MRSigna2247157
TPS Reset Fails Due to RRF Exciter.
Symptom
Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.
Error message
2247157 The Master Exciter in the RRF chassis (rear, slot 7) was not detected. Error: 2247157 Host: ge3T Process: diagCoord File: Rrf.cpp Line: 249 Task ID: 0x3f1fdf8 Time: 999 Object: APS The Master Exciter in the RRF chassis (rear, slot 7) was not detected. The system will not function without this board.-> Error: 2247157 Host: ge3T Process: diagCoord File: Rrf.cpp Line: 249 Task ID: 0x3f1fdf8 Time: 999 Object: APS The Master Exciter in the RRF chassis (rear, slot 7) was not detected. The system will not function without this board. Extended Error: The master exciter provides the main clock for the RRF chassis and the IRF board. Verify that the board is powered and that its LO DETECT LED is on. If this is a 16 channel system, verify that the two fiber cables on the two DIF boards are not switched with each other. Possible cause(s): IRF Board, Exciter Board, RRF-DIF Board, GigaLink Fiber Cable
Applies to
- Models
- Signa
- Product
- Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
- Software
- 15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
- Component
- 2361470 or 2361470-2 REMOTE RF MASTER EXCITER BOARD or RRF Master Exciter 2 Board with Temperature Compensation 2361470-3 2361470 or 2361470-2 REMOTE RF MASTER EXCITER BOARD or RRF Master Exciter 2 Board with Temperature Compensation 2361470-3
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.