Pular para o conteúdo principal
GE Healthcare · CTLightSpeed26000777426000740250444

LightSpeed 5.X (60kW)

Symptom

Intermittent DAS communication errors. Problem occurs if scanning is paused or aborted. See Long Description Once fault occurs the exam needs to be ended and a scan hardware reset performed at a minimum to resume scanning again.

Error message

Error: 260007774 List of Converter Cards From the Previous Error Message Card List: xx xx See Long Description for full error Tue Dec 30 09:25:17 2008 Error: 260007774 Host: orp Process: orp_subsys File: CnvBoardInterface.cc Line: 852 Tue Dec 30 09:25:17 2008 Host : Das Ermes # : 260007774 Exception Class : Secondary Severity : Sec/Soft File : CnvBoardInterface.cc Line# : 852 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 List of Converter Cards From the Previous Error Message Card List: 8,11,14-15,19,21,25-27,29-36,38-58,60-63 Note on this error. Sometimes the actual bad board will be the one that is not listed.

Error: 260007402 Converter Board(s) Asserted a Board FAULT. See Long Description for complete error Tue Dec 28 09:41:31 2004 Error: 260007402 Host: orp Process: orp_subsys File: cnvmgr.c Line: 605 Tue Dec 28 09:41:30 2004 Host : Das Ermes # : 260007402 Exception Class : Abort Severity : Secondary File : cnvmgr.c Line# : 605 Function : No System Function Reported Scan Type : Static Scan: 50444/1/1 Converter Board(s) Asserted a Board FAULT.

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues