GE Healthcare · CTHiSpeedVCT26000741620053183211
Scanner Hardware Stopped Scan message during axial scan
Symptom
Scanner Hardware Stopped Scan message during axial scan. Long Description for Feedback Only. Scanner Hardware Stopped Scan message during axial scan
Error message
Error: 260007416 Detector Temperature Is High. Image quality may be affected See Long Description Tue Sep 6 09:07:45 2005 Error: 260007416 Host: orp Process: orp_subsys File: dtemptst.c Line: 364 Tue Sep 6 09:07:44 2005 Host : Das Ermes # : 260007416 Exception Class : Soft Severity : Sec/Soft File : dtemptst.c Line# : 364 Function : No System Function Reported Scan Type : Scout Scan: 372/1/1 Detector Temperature Is High. Image quality may be affected. Current Reading:40.41 Deg C A/D Counts:20053 Error 183211 OBC Ambient maximim temperature exceeded See long description for full error Mon May 5 2008 07:53:04 Error 183211 Host OBC OBC Ambient maximim temperature exceeded. Actual temperature = 54.58 (C) Temperature limit = 54.99 (C)
Applies to
- Models
- HiSpeed, VCT
- Product
- HiSpeed QX/i - GRE Always MDAS 4 QX_P_M4 HiSpeed QX/i - GRE Method of visual identification Small rounded gantry similar in appearance to LS 3.X and 4.X. Always has an MDAS 4 slice Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. xxx
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required