GE Healthcare · CTLightSpeedVCT260111601
LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry.
Symptom
Step and Shoot (SmartStep or SmartView) hand controller not functioning See LD for more Step and Shoot (SmartStep or SmartView) hand controller not functioning When customer attempts to use the hand controller for VCT step and Shoot, the system will not initialize into step and shoot mode.
Error message
Error: 260111601 IHHC reset timeout. See LD for complete error Thu Oct 20 12:41:32 2005 Host : Table Gantry Processor Ermes # : 260111601 Exception Class : Abort Severity : Primary File : hc_exec.c Line# : 766 Function : No System Function Reported Scan Type : None/Unknown Scan: 4628/4/0 IHHC reset timeout.
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed RT16 & RT16 Xtra and Vision WB_H_G16_G_GTL = RT16, Xtra WB_H_G16_G_GTL650 = Vision WB_H_G16_G_GTM = RT16, Xtra or Vision LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry Links to the Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5370000 - TGPHD BOARD ASSY - TGPU For HD, HP and VCT (Replaces the 5271005-2 & 2349697-2, -3, -4 and -5). See Long Description about backward compatability. Part #: 5271005-2 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: TGPHD BOARD ASSY Positioning VCTHD FRU: Yes Status: Orderable Comments: Per ECR 2098435, 212008898, 30 April 2010. Old FRU Compatible Yes Comments Per ECR 2098435, 212008898, 21 Jul 2010. Additional FRU Needed No Old FRU Scrapped No Old FRU Upgradable No Use on Hands Yes Use Demand History Yes Ultimate Prime 5271005-2 New FRU Backward Compatible Yes Old FRU Usable Yes Change Type Indirect ECO Number ECR2098435 2349697-5 is replacing the -4 version of this board. -3 versions are still in stock. If you are replacing parts in a VCT or HP gantry order the -3 until stock is depleted. ** FE ASSIST ** The 2349697-3 version is for VCT and all 5.x systems i.e. HP60 HP80- HP100's. The 2349697-4 version will works for CT750-HD and is backwards compatible but the -3 will not work on a CT750HD. (212001261 12/11/08) Part #: 2349697-5 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: TGPU BD ASSY Positioning VCTHD FRU: Yes Status: Orderable Part #: 2349697-4 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: TGPU BD ASSY Positioning HD FRU: Yes Status: Orderable Comments: ** FE ASSIST ** The 2349697-4 is for CT750-HD and is backwards compatible to The 2349697-3 used on VCT and all 5.x systems i.e. HP60 HP80- HP100's. Do not use the 2349697-3 on CT750HD (212001261 12/11/08) TGPU 2349697-3 Part #: 2349697-3 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: TGPU BD ASSY Positioning HP60 FRU: Yes Status: Orderable Comments: Added per VCT RSPL, this is replacement for -2, and non dash versions. ms13452 050325
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required