Pular para o conteúdo principal
GE Healthcare · MRSigna2277651

HDx TPS Reset Failure Due to a Image Compute Node ICN.

Symptom

Unable to Download or Reset the TPS MGD or ISE Unable to download TPS or ISE or MGD chassis

Error message

2277651 ICN Hard Disk Health Diagnostic failed. Mon Feb 23 06:44:11 2009 Error: 2277651 Host: EPMR02 Process: diagCoord File: icnDiskHealth.pl Line: 363 Exception Class: Notification Function: Software/Firmware/Config: N/A ICN Hard Disk Health Diagnostic failed. icn1 - No disks to process->

Applies to

Models
Signa
Product
3.0T Signa HDxT For 3.0T Signa HDxT systems with 15.0 software only. For 3.0T Signa HDxT systems with 15.0 software only.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
SecondOpinion Image Compute Node 5127452 See comments in Resolution Concept for second opinion requirements before ordering this part.
Area
MR

Full cause & step-by-step resolution