GE Healthcare · MRSigna2252294
Cable found shorted between driver module and rear IF panel. (problem occured during HD to HDxt upgrade).
Symptom
Unable To Scan With Driver Module Errors Unable To Scan With Driver Module Errors.
Error message
2252294 The Driver Module Has Detected an Invalid Configuration. Source of Fault is Multi-Coil Board #2 Not Present See Long Description 2252294 Host: sfhmr1 Process: NSP File: SCP:DCB Line: 283 The Driver Module Has Detected an Invalid Configuration. Source of Fault is Multi-Coil Board #2 Not Present.-> Extended Error: The requested Driver configuration has been determined to be invalid. The source of the fault will give specific information regarding the fault. Fault could be caused by hardware or software
Applies to
- Models
- Signa
- Product
- Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
- Software
- 15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.