Legacy No power to the table or IDD assembly

Symptom

No power to the table or IDD assembly

Error message

Error 4B58 on console See long description for details on error Error 4B58 - Intelligent Spot Film Device function not availableDebug message: RF: Lost ISFD linkRecommended actions:1. Inspect the cable harnesses connectin gthe ISFD DSC to the R&F positioner, including those cables connecting the card rack to the patch panel. Secure or replace accordingly.2. Inspect the LEDs on the side of the ISFD DSC FRU. These may indicate a boot failure of the ISFD firmware. In whcih case, replace the ISFD DSC.3. Run system level diagnostics on the R&F Interface board. Replace the board of necessary.4. Replace the ISFD DSC FRU.

Full cause & step-by-step resolution