GE Healthcare · MRSigna2249436
VRE Reset Failure Due to ICN.
Symptom
Unable to Download or Reset the TPS MGD or ISE Unable to download TPS or ISE or MGD chassis
Error message
2249436 Communication problem between APS and VRE. See long description Mon Jun 4 07:37:51 2007 Error: 2249436 Host: lx-mr Process: diagCoord File: cm_init.c Line: 293 Task ID: 0x6d89800 Time: 1807 Object: APS Exception Class: Unknown Function: Software/Firmware: Host Communications Communication problem between APS and VRE. The APS could not notify the VRE that communication is established.-> Extended Error: ERROR: EM_OS_CM_INIT_RPC_VRE Communication problem between APS and VRE. The APS could not notify the VRE that communication is established. Extended Error: This could be an indication of a hardware problem with the APS or VRE boards or an ethernet cabling problem on the MGD subnet. Verify the ethernet cabling on the MGD subnet, reboot VRE, and perform a TPS Reset. Run diagnostics if the problem recurs.
Applies to
- Models
- Signa
- Product
- Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
- Software
- 15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
- Component
- SecondOpinion Image Compute Node 5127452 See comments in Resolution Concept for second opinion requirements before ordering this part.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.