Processor Exception Bus Errors and not enough memory available to start acquisition errors due to AP memory.
Symptom
System Hangs During Image Reconstruction
Error message
2227700 The checksum of the vxWorks platform executable code has changed Fri Oct 23 14:36:13 2009 Error: 2227700Host: lx-mr Process: NSPFile: TPS:lo check daemon: system TEXT er Line: 0The checksum of the vxWorks platform executable code has changed. 2232902 Processor Exception Frame, Bus Error: vector 2, interrupt level 0 Mon Nov 3 13:11:30 1997 Error: 2232902 Host: oc001 Process: NSP File: TPS:excCHECKER Line: 0 Processor Exception Frame, Bus Error: vector 2, interrupt level 0 current pc 0x279bf2, sr 0x3004, task name MC860_CEID_2, fault address 0x400b0008, fault status 0x105 additional information in /usr/g/service/log/vxExLg-4c752740 2218527 Scanning has stopped. not enough memory available to start acquisition, please try again. 2218527 Host: MR_MT305 Process: scn File: MsaScanOpsScrn.m Line: 909 Scanning has stopped. not enough memory available to start acquisition, please try again. Error: 2218527 Host: MR_MT305 Process: scn File: MsaScanOpsScrn.m Line: 909 Scanning has stopped. not enough memory available to start acquisition, please try again.
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Component
- SecondOpinion 128MB PMOD Memory Board 2292662-X This part requires second opinion approval before ordering. See comments in resolution concept for second opinion requirements before ordering this part. This part requires second opinion approval before ordering. See comments in resolution concept for second opinion requirements before ordering this part. - The PMOD board, part number 2292662-3 is compatible with the 2292662-x and 2292662-1x Medcam baseboards. - The PMOD board, part number 2292662-3 is NOT compatible with the 2101767 and 2222532 Medcam baseboards (this includes 2101767-16, -18, -20, and –22. 2222532, -4, -5, and –7). These Medcam baseboard models require the 128 Mb MOD memory board part number 2101767-5. - Mixing this PMOD board 2292662-3 with the old MOD 128MB Memory board, part # 2101767-4, is NOT supported.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- CWA ID100315. Auto and Manual Prescan Fails.
- What is the Acoustic Noise Level in the Ovation Magnet.