Pular para o conteúdo principal
GE Healthcare · MRSigna

ICN configuration failed With Waiting 1200 (max) seconds for ICN OS load/reboot Due to ICN.

Symptom

ICN configuration failed With Waiting 1200 (max) seconds for ICN OS load/reboot : watch_logfile: only found 0 of expected 2 systems before timeout Error ICN configuration failed With Waiting 1200 (max) seconds for ICN OS load/reboot : watch_logfile: only found 0 of expected 2 systems before timeout Error.

Error message

ICN configuration fails with message watch_logfile, only found 0 of expected 2 systems before timeout see long for entire message watch_logfile: only found 0 of expected 2 systems before timeout DHCP IP/MAC list: FAILURE: The ICN OS load has exceeded the timelimit. OPSTAT: complete_error STATE: restart ERROR: ERROR: DHCP address requests failed.Only received 0 of 2 DHCP requests after OS install. Please check ethernet cables, reseat connections and retry VRE configuration.Exiting VRE configuration. FAILURE: ERROR: ICN Install. The load operation has timed out. VRE configuration has failed. Exiting.

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
SecondOpinion Image Compute Node 5127452 See comments in Resolution Concept for second opinion requirements before ordering this part.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.