Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT2601000592836026011100226011100311569244634

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

No lights on gantry. Customer was unable to scan or position table, all table gantry hardware was off-line No lights on gantry. Customer was unable to scan or position table, all table gantry hardware was off-line

Error message

Error: 260100059 DIP Ready-For-Views message received before offset triggers were stopped. See LD for Entire Error. Wed May 2 14:02:34 2007 Error: 260100059 Host: tgp Process: tgp_subsys File: scTrigSeqActions.c Line: 1927 Wed May 2 14:02:32 2007 Host : Table Gantry Processor Ermes # : 260100059 Exception Class : Abort Severity : Primary File : scTrigSeqActions.c Line# : 1927 Function : No System Function Reported Scan Type : Helical Scan: 28360/4/1 DIP Ready-For-Views message received before offset triggers were stopped. HW/SW Error: Probable problem with das trigger or das view hardware or cables. Possible sofware sequencing problem.

Error: 260111002 GUIF CAN bus error warning See the LD for the full error code Error: 260111002 GUIF CAN bus error warning Mon May 16 07:45:39 2005 Error: 260111002 Host: tgp Process: tgp subsys File: guif can mgr.c Line: 483 Mon May 16 07:45:39 2005 Host : Table Gantry Processor Ermes # : 260111002 Exception Class : Soft Severity : Sec/Soft File : guif can mgr.c Line# : 483 Function : OPERATOR INPUT/OUTPUT : Gantry UIF Communications Scan Type : None/Unknown Scan: 0/0/0 GUIF CAN bus error warning. 96 CAN bus errors have occured. GUIF CAN bus error 

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
U8009FF Fuse 3.2 Amp 250v See LD for details Part #: U8009FF Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: FUSE FRU: Yes Status: Orderable
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues