GE Healthcare · MRSigna225363319993622480082253614

IRF RRF data link type errors Due to poor connection of gigalink cable from J21 of RF-DIF(slot 10-11 RRF chassis) to J20…

IRF RRF data link type errors Due to poor connection of gigalink cable from J21 of RF-DIF(slot 10-11 RRF chassis) to J20 IRF I/O(slot 11 MGD chassis).

Symptom

Scans Stop Intermittently With Error Code 2248008 "RFM EOS semaphore timeout error" See Long Description Scans stop in the middle of the scan with the following error logged. This happens about once per day. (PQR 13096684) 2248008 EXCITE15T NSP APS:DRF 250 RFM EOS semaphore timeout error. IRF did not send enough data to DRF. Please change prescription or perform a TPS Reset. DRF ID: 1 FE ID: 1 segCount: 12 frameCount: 14

Error message

2253633 Link errors were detected on IRF RRF data link GB1 see long Error: 2253633 Host: MRS2 Process: diagCoord File: Irf.cpp Line: 2176 Task ID: 0x26aac90 Time: 199936 Object: APS Exception Class: Unknown Function: Software/Firmware/Config: N/A Link errors were detected on IRF<->RRF data link GB1. RX signal ready loss : 0 (1 = Yes, 0 = No) Irf Error Frame : 1 (1 = Yes, 0 = No) RRF Error Frame : 0 (1 = Yes, 0 = No)->

2248008 RFM EOS semaphore timeout error. IRF did not send enough data to DRF.

2253614 Diags detected checksum errors on the IRF RRF data link GB1 see long Error: 2253614 Host: MRS2 Process: diagCoord File: Irf.cpp Line: 2145 Task ID: 0x26aac90 Time: 199936 Object: APS Exception Class: Unknown Function: Software/Firmware/Config: N/A Diags detected checksum errors on the IRF<->RRF data link GB1. Irf->Rrf chksum errs: 0, Irf->Rrf missed counts: 0 Rrf->Irf chksum errs: 0, Rrf->Irf missed counts: 271->

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required