Pular para o conteúdo principal
GE Healthcare · MRSigna2248830

2248830 Analogic RF Amplifier 2 Power Loss Fault due to RF Amplifier.

Symptom

Will Not Scan With RF Errors Will not scan with RF errors

Error message

2248830 Analogic RF Amplifier 2 Power Loss Fault Sat Oct 24 10:39:40 2009 Error: 2248830Host: UWMR3 Process: NSPFile: SCP:../RfNbMdsHWO.cpp Line: 2505 Exception Class: Unknown Severity: UnknownFunction: RF: RF TransmitPSD: FSE-XL Coil: 8HRBRAIN Scan: 4331/9Analogic RF Amplifier 2 Power Loss Fault.RF Interface (RFI) detects no power to Analogic RF Amplifier 2.Possible cause: the breaker on the front of the RFI is turned off.-

Applies to

Models
Signa
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
2224562-2 1.5T RF AMP POWER MODULE W/ FRU PKG (SRFD)