Pular para o conteúdo principal
GE Healthcare · CTLightSpeed260007384

LightSpeed 4.X - GRE

Symptom

Scan hardware stopped scan with RCIB errors

Error message

Error: 260007384 Scan ABORT -- DAS Has Detected An Error.^C The DAS has received an out-of-sequence command. Command 3000 cannot be accepted when the DAS is in the State: Normal State and Command Level: Scan in Progress Sat Dec 11 13:17:07 2004 Error: 260007384 Host: orp Process: orp_subsys File: invalcmd.c Line: 141 Sat Dec 11 13:17:06 2004 Host : Das Ermes # : 260007384 Exception Class : Abort Severity : Secondary File : invalcmd.c Line# : 141 Function : No System Function Reported Scan Type : Helical Scan: 1609/1/1 Scan ABORT -- DAS Has Detected An Error.^C The DAS has received an out-of-sequence command. Command 3000 cannot be accepted when the DAS is in the State: Normal State and Command Level: Scan in Progress.

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2159578-3 collimator control board used on 1X/2X/3X/4X/5X See Long description for details Part #: 2159578-3 Part Type: GP CONSUMABLES Part Weight: 2.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: COLLIMATOR CONTROL BOARD FRU: Yes Status: Orderable Comments: Not Available
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues