GE Healthcare · CTLightSpeedVCT260134816118634500001000004759482943
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
Cradle does not move on GT table or stuck at the home position. Table moves OK, but Cradle will not drive IN or OUT of gantry.
Error message
Error: 260134816 IMS velocity is out of tolerance. See Long Description for a complete error Fri Sep 22 09:51:38 2006 Error: 260134816 Host: tgp Process: tgp_subsys File: gtcb_velocity_chk.cxx Line: 295 Fri Sep 22 09:51:38 2006 Host : Table Sub-System Ermes # : 260134816 Exception Class : Secondary Severity : Pri/Soft File : gtcb_velocity_chk.cxx Line# : 295 Function : No System Function Reported Scan Type : None/Unknown Scan: 6856/5/0 IMS velocity is out of tolerance. current pos (um) : 118634 velocity tole(um/s) : 50000 expected vel (um/s) : 100000 current vel (um/s) : 47594 start pos (um) : -58 samples:32 at spd loc:26 cntr:50 pos[at spd loc]:82943
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 5133506-2 RH Top Side Cover Used on VCT GT1700 table (replaces 5133506) See Long Description for details and picture 5133506-2 RH Top Side Cover VCT GT 1700 replaces 5133506 Part #: 5133506-2 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: TOP SIDE COVER R 1700 ASSY Positioning GT FRU: Yes Status: Orderable Comments: added per to ECR2052664 by sso#305010201 in Dec 10, 2008 Part #: 5133506Part Type: GP CONSUMABLES Part Weight: 0.00 LBRRepairable: NoReturnable: NoHazardous Material: NDescription: TOP SIDE COVER R-1700 ASSY positioning GTFRU: YesStatus: Orderable
- Area
- CT
Full cause & step-by-step resolution