Symptom

Table is lowering on its own with Uncommanded elevation motion. motor controller detects uncommanded motion. - document in RFS - critical issue Table is lowering on its own with Uncommanded elevation motion. motor controller detects uncommanded motion. Problem is becoming less intermittent and exam sometimes proceeds normally once the resume button is selected. The problem also occurs more frequently when patient is heavy. ct01% /usr/g/insite/bin/viewlog Tue Aug 10 19:04:09 2004 Error: 184836 Host: ETC Process: Elevation File: elev scan.c 1.17 9/30/92 Line: 178 Function: No System Function Reported Scan: 4530/2/7 Type: Helical Exception Level: Pri/Most Severe Time: 19:04:08:905 Log Series: 376 motor controller detects uncommanded motion. Aborting any active scan

Error message

Error: 260115209 Elevation stopped at invalid position. see long description for details Sat Mar 15 06:13:29 2008 Error: 260115209 Host: tgp Process: tgp subsys File: mtcb emc motion check.c Line: 294 Sat Mar 15 06:13:30 2008 Host : Table Sub-System Ermes # : 260115209 Exception Class : Secondary Severity : Pri/Soft File : mtcb emc motion check.c Line# : 294 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Elevation stopped at invalid position. expected target position = 515825 um actual stopped position = 516693 um

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2343479-2 Table Elevation Linear Actuator , H-PWR 5.x Table See Long Description for details Part #: 2343479-2 Part Type: GP CONSUMABLES Part Weight: -98999.01 LBRRepairable: NoReturnable: NoHazardous Material: NDescription: Actuator, Linear, H-Power based Tables, PinnedFRU: Not AvailableStatus: Orderable Comments: Added per ECR 2007660 as replacement for 2342479. ms13452 041217
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.