Problems with Dicom images sent to CMS Treatment Planning System.

Symptom

Problems with Dicom Images Sent to CMS Treatment Planning System. Images are sending with no issues reported in the log at this destination. This unit is used prior to a radiation therapy session. MR and CT images are merged and used to determine the area of interest to be treated. This issue the CMS unit is having is with the image format when it shows up on the system database. File names include a "-" (dash). The system will not accept this.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR