LightSpeed 4.X - Octane

Symptom

Soft ring artifact appears on all scans with no errors in the error log. See LD for details Soft ring artifact appears on all scans with no errors in the error log. The recon selected was BONEPLUS. Note that RINGFIX is not applied to bone or detail algorithms. This is the cause of why the soft tissue looks very noisy. This is normal for this recon as the window and level would generally be adjusted to view the bone - not the soft tissue.

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 4.X - Octane sw_version: 405I.2_H4.0M5 LightSpeed 4.X - Octane Gantry Model # 2339985 Console Model #'s 2341923 and 2341923-2(D3D) check_config shows LightSpeed 16 / LightSpeed 16 Product: LightSpeed16 sw_version: 405I.2_H4.0M5 or newer machine_type: Octane host_id: 3232297858 tube_type: 15 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution