Symptom
Scanner Hardware Stopped Scan with DAS parity errors
Error message
Error: 260007386 DAS Has Detected An Error. Data Parity Error detected in the xx Data Stream(s) on Converter Board xx or the DCB Board. See LD for sample errors Wed Mar 23 15:46:11 2005 Error: 260007386 Host: DAS Process: DCB Firmware File: parity.c Line: 219 Function: DATA ACQUISITION : Scan Data Acquisition Scan: 214/5/1 Type: Helical Exception Level: Sec/Soft Time: 15:46:11:545 Log Series: 660 SCAN ABORT -- DAS Has Detected An Error. Data Parity Error detected in the 1B Data Stream (s) on Converter Board 33 or on Converter Board 41 . NOTE: Before troubleshooting, go through gesys.log (viewlog) and search for all “Data Parity Error” reports in the log. There are 2 items you need to note from the error: 1- Data Parity Error detected in the xx Data Stream and 2- On Converter Board xx or on Converter Board xx Tue Jan 10 09:37:14 2006 Error: 260007386 Host: orp Process: orp_subsys File: parity_16.c Line: 255 Exception Class : Abort Severity : Secondary Function : No System Function Reported Scan Type : Static Scan: 50056/1/1 SCAN ABORT -- DAS Has Detected An Error. Data Parity Error detected in the 3A Data Stream (s) on Converter Board 55 or on Converter Board 71 . Note: Error 260007
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 2.X (Plus, QX/i) - Octane Always S-DAS 2.4.2_H2.4M5 LightSpeed 2.X (Plus, QX/i) - Octane (Pegasus or Rhapsode IG) - Always S-DAS Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 2.4.2_H2.4M5 FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 FEATURES/OPTIONS STD/OPTION Cat # Digital Laser Camera I/F B7700PD Smart Helical (s/w only) Standard B7530TK Fast Recon (s/w only) Option B7700JF BMD (requires B7710MP or MN) Option B7710MK Smart Prep Option B7530SP Smart Score ? ? 3D Analysis (Requires D3D) Option B7540TD Navigator (s/w only) Option B7540N DentaScan (s/w only) Option B7540SD Direct 3D (Requires B7710SA) Option B7810SS HIS/RIS Option (h/w and s/w) Option B7500PN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2137958-4 SDAS Convertor Board (replaces 2137958-3) May use 2137958-3U converter Board if no current stock of the 2137958-4 See long description for details Part #: 2137958-4 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: -39999.60 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: CONVERTOR WITH BLACK BEUTY CONNECTOR FRU: Yes Status: Orderable Comments: OBSOLETE PART case 10788898 This part has been identified as obsolete and no short-term solution has been identified FE must identify and order a higher or lower level spare part as required Contact your support engineer on-line engineer or Parts ID team should you require support 212012173 06-26-07 REPLACES P/N 2137958-3 THE NEW -4 IS COMPATIBLE TO THE -3. MS20544 ON 3/7/03. ** FE ASSIST ** May use the USED part 2137958-3U converter Board if no current stock of the 2137958-4. (9/10/2007 212001261)
- Area
- CT
Full cause & step-by-step resolution