Cannot Boot Applications with Host Monitor Timeout Error due to Configuration Files and Software are Corrupt due to Powe…

Cannot Boot Applications with Host Monitor Timeout Error due to Configuration Files and Software are Corrupt due to Power Outages at Site.

Symptom

Cannot Boot Applications with Host Monitor Timeout Error

Error message

2248349 The file GradientConfig.cfg contains a missing or bad value. See Long Description This error may indicate one of several different data values that are corrupt. Some examples are: GCoilType SRMode

2248349 The file MRconfig.cfg contains a missing or bad value. See Long Description This error will include one of many different data values that can be corrupted. Some examples are: begTravel boreTempHyst boreTempLev1 boreTempLev2 boreTempOffset digiTuningBoard endTravel fieldStrength gradientAmp gradProcessorType isoVectorZ lineFreq mCoilTRdriver PowerMonType realTimeMon RFampType rrfCalibFlag specRFamp spectroRFampType usssType

Hostmonitor Timeout after 100 seconds

Applies to

Models
Signa
Product
3.0T Signa HDxT For 3.0T Signa HDxT systems with 15.0 software only. For 3.0T Signa HDxT systems with 15.0 software only.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Area
MR