Pular para o conteúdo principal
GE Healthcare · CTBrightSpeedLightSpeed260110005

BrightSpeed HPower or Delight HPower

Symptom

Scan aborts WITH Communications with the CCB sub-system has not been established. Scan aborts WITH Communications with the CCB sub-system has not been established

Error message

Error: 260110005 The DAS/collimator firmware object had a timeout waiting for message 3003 See long description for full errors Sat Mar 21 19:14:02 2009 260110005 orp orp_subsysdas_event_mgr.cxx 1884Sat Mar 21 19:14:01 2009Host : On-Board Rotating Processor Ermes # : 260110005Exception Class : Secondary Severity : Pri/SoftFile : das_event_mgr.cxx Line# : 1884Function : No System Function ReportedScan Type : None/Unknown Scan: 0/0/0 The DAS firmware object had a timeout waiting for message 18 Sat Mar 21 19:14:02 2009 260110005 5 orp orp_subsyscoll_event_mgr.c 1357Sat Mar 21 19:14:01 2009Host : On-Board Rotating Processor Ermes # : 260110005Exception Class : Secondary Severity : Pri/SoftFile : coll_event_mgr.c Line# : 1357Function : No System Function ReportedScan Type : None/Unknown Scan: 0/0/0 The Collimator firmware object had a timeout waiting for message 18

Applies to

Models
BrightSpeed, LightSpeed
Product
BrightSpeed HPower, Delight, Velocity or MERI HPower HP_P_G4_G_HPT Excel HP_P_G8_G_HPT Edge HP_P_G16_G_HPT Elite HP_P_G16_G_GMV Elite HP_P_P16_G_GMV Elite HP_P_M16_G_HPT BrightSpeed Method of remote software hardware identification check_config swhwinfo List of Supported Console Types : HP_P_G16_ G _HPT Note: G could be GRE, AIO or TIO - check with: cat /usr/g/config/INFO | grep CONSOLETYPE GOC4 AIO TIO List of softwares the product can run: 08BW20.1 07MW11.10 08BW44.1 ( Delight ) 09HW30.4 ( Velocity ) 09BW35.11 ( MERI - Pancake DAS ) FAQ's 1. All BrightSpeed Power Systems use the Performix tube D3186T HP_P_G4_G_HPT (Excel - H-Power Table) HP_P_G8_G_HPT (Edge -H-Power Table ) HP_P_G16_G_HPT (Elite -H-Power Table ) HP_P_G16_G_GMV (Elite -Global Table Mid Value - modified GT1700 no IMS ) Note: G could be GRE, AIO or TIO (check with: cat /usr/g/config/INFO | grep CONSOLETYPE ) BrightSpeed Elite Power Systems will only be sold as Velocity System starting in 2Q/2010. Delight Elite Power being phased out 1Q/2010 BrightSpeed Power 4 & 8 slice systems have been discontinued due to restructuring of the product lineup. Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2159578-3 collimator control board used on 1X/2X/3X/4X/5X See Long description for details Part #: 2159578-3 Part Type: GP CONSUMABLES Part Weight: 2.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: COLLIMATOR CONTROL BOARD FRU: Yes Status: Orderable Comments: Not Available
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues