Defective IPG bd as identified by diagnostic failure in walking 1s and 0s test

Symptom

Auto prescan Fails with Signal Too Small message on all coils Auto prescan fails with Signal Too Small message on all coils

Error message

2224666 Auto Prescan failed, signal too small. Please try Manual Prescan. Fri Jan 23 14:39:55 2004 2224666 lx-mr psc PscMainControl.cc 1271 Auto Prescan failed, signal too small. Please try Manual Prescan.

2224580 Auto Prescan failed. Signal was too small to process - use Manual Prescan 2224580 Host: hopgemr1a Process: NSP File: APS:aps_cf Line: 11 Auto Prescan failed. Signal was too small to process - use Manual Prescan

2222267 IPG Sequencer RAM walking 1s and 0s test failed. see long description Error seen with TPS diags: 1094647477 1 1 Wed Sep 8 12:44:37 2004 2222267 2 lx-mr diags diagsMain.cc 2263 IPG BOARD LEVEL DIAGS FAILURE IPG Sequencer RAM walking 1s and 0s test failed. Pattern Written: 00000000H Pattern Expected: 00000000H Pattern Read: 00000080H Address: 00d80002H EN 291 SR 310

2224709 Low signal detected. Please check the signal level for receiver x or receiver x+1 while using Manual Prescan. Error: 2224709 Host: NMI_Langhorn Process: NSP File: TPS:aps_frame_range_check Line: 10 Low signal detected. Please check the signal level for receiver 1 (or receiver 2 while using Manual Prescan). Note that for 3 coil scans, one coil will cause a low signal detected message to 

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR