GE Healthcare · CTBrightSpeedLightSpeed26010102350444260007402260007453260001512260007353
BrightSpeed Select or Delight Select
Symptom
"Scan hardware Stopped Scan" with collimator errors. "Scan hardware Stopped Scan" with collimator errors
Error message
Do Not Use - use on of the other two Error: 260101023 The (DCB or CCB) has notified the system of an error see long description for examples. Tue Dec 28 09:41:31 2004 Error: 260101023 Host: orp Process: orp_subsys File: rcib_erlog.c Line: 366 Tue Dec 28 09:41:30 2004 Host : Das Ermes # : 260101023 Exception Class : Abort Severity : Primary File : rcib_erlog.c Line# : 366 Function : No System Function Reported Scan Type : Static Scan: 50444/1/1 The DCB has notified the system of an error condition which prevents scanning. See log for further information:^C Error message code: 260007402 Notification level: 5 Sequence number: 1 The above error is a catch all for many different notifications and sequences. Other examples: Host : Das Ermes # : 260101023 Exception Class : No System Function Reported Scan Type : Helical Scan: 8564/2/1 The DCB has notified the system of an error condition which prevents scanning. See log for further information: Error message code: 260007453 Notification level: 5 Sequence number: 17 Wed Sep 22 11:21:43 2004 Error: 260101023 orp orp_subsys rcib_erlog.c 363 Wed Sep 22 11:21:41 2004 Host : Das Ermes # : 260101023 Exception Class : Abort Severity : Primary Fil
Applies to
- Models
- BrightSpeed, LightSpeed
- Product
- BrightSpeed Select, Delight Select or MERI Select Elite-16, Edge-8, Excel-4 HP_S_G16_G_HLT HP_S_G8_G_HLT HP_S_G4_G_HLT BrightSpeed Select or Delight Select (Hivalue Lite) 4/8/16 Slice Scanner BrightSpeed Delight Select Systems started to ship in first Quarter or 2009 with AIO (All in One) Console. AIO console uses HP8400 and NO DARC or IG's. Recon time is 6 FPS. If customer orders Fluoro option system requires one IG. The 16 FPS Upgrade in Not Available for Select systems. BrightSpeed Meri Select Systems ( Pancake Das ) started to ship in 3rd quarter of 2010, 4/8/16 Slice Scanner HiValue 16 Lite −−> BrightSpeed Elite Select HiValue 8 Lite−−> BrightSpeed Edge Select HiValue 4 Lite−−> BrightSpeed Excel Select BrightSpeed Select Scope: - 3.5MHu Solarix tube - JEDI 60 dc Lite - Modified NP Table Key Features: - 16x0.625mm(16x1.25) for Lite 16 - 8x1.25/8x2.5mm for Lite 4/8 - Rotation Speed: 0.8 s - Gantry: 2040mm width, 70cm Bore - Table: 1337@ISO-140mm scannable range - Recon: 3fps (w/o IG @ 2G Memory(4/8), 4G Memory(16)), optional: 6fps(w/ 1IG ) - Power: 42KW JEDI power See the attached Service Intro Planner for details Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2159578-3 collimator control board used on 1X/2X/3X/4X/5X See Long description for details Part #: 2159578-3 Part Type: GP CONSUMABLES Part Weight: 2.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: COLLIMATOR CONTROL BOARD FRU: Yes Status: Orderable Comments: Not Available
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.