GE Healthcare · CTBrightSpeedHiSpeed186218

CT/e, CT Prospeed AI/FI Irix or Linux

Symptom

Scan Aborts with JEDI XG Error. Scout Scan Aborts with JEDI XG Error

Error message

Error:186218 Detail error code:50 - 0204H Heater Error: Open circuit (HW detected) See long description for complete error Host: IRIS Proc:tgp_out Error:186218 File:tgp_in_input.c Line:334 XG Error phase is 3 code is 50 Detail error code:50 - 0204H Heater Error: Open circuit (HW detected)

Applies to

Models
BrightSpeed, HiSpeed
Product
CT/e, CT Prospeed AI/FI Irix or Linux See long description for sale information Product Name: CT/e Single CT Scanner: Irix Platform A list of types. (a Product/hardware table) N/A An On-site / Remote identification aid . (Example: check_config) N/A A Table of options (included, optional, new introductions, …) N/A A list of Software that has run on this product (Product / Software Compatibility Table) 3.01 / 3.02 / 3.03 Linux CTe Product photographs N/A Marketing Information N/A Documents: Always use the most appropriate documentation for your system. Documentation can be found on CT Service CDROM/DVD, or online at FCT Service Engineering Tech Pubs Website, Common Document Library, and My Workshop. Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found in the attachment Always use the most appropriate documentation for your system.
Component
2214527 Jedi standard HV tank for HiSpeed Xi, HP 60 and BrightSpeed See Long Description for details Part #: 2214527 Part Type: GP CONSUMABLES Part Weight: -3059969.40 LBR Repairable: No Returnable: No Hazardous Material: N Description: JEDI STD HV TANK (SPARE) FRU: Yes Status: Orderable Comments: FLS - JEDI STD HV TANK U/O YMS PROSPEED LXI,FXI,DXI REPLACES 2186730 DIFFERENCE IS PACKAGING FOR SERVICE. HQT: QUALITY ISSUE 10/98 ALL TANKS WITH SERIAL NUMBER PRIOR TO 22124BU4 ARE BAD. FE SHOULD CHECK HIS SERIAL # EUROPE PURGING GLOBE WILL SEND NEW MATERIAL AS QUICKLY AS POSSIBLE. AMERICA HAS NO STOCK ONHAND AS OF 10/22/98MS08992 RETURN DEFECTIVES TO: SWAP (MICHELL CHRISTOFFERSON), 2200 E. COLLEGE, BUILDING 11, CUDAHY, WI 53110. ADDED BY MS20544 ON 3/25/02. RETURN DEFECTIVES TO RECYCLING7/23/01 MS5277 TAKEN OFF RECYCLING PROGRAM 8/30/01 MS5277 TAKEN OFF RECYCLING PROGRAM PER L.BENIWAL (MSA8429 09/05/02) GELS - 2214527 = 2186730 + WOODEN BOX 10DISPUTES DEFECTIVE PARTS TO BE PUT ON SHELVES FOR ANALYSIS PDB00/07/27 KANBAN CSS F00,JEDI STD HV TANK (SPARE) 2214527 Jedi standard HV tank for HiSpeed and HP 60 Systems HV Tank for HP60 systems - must be disassembled from the Inverter Power assembly, and the new tank installed. Make certain you follow the disassembly/reassembly instructions in the Service Manuals exactly, as the process is quite difficult, and you are prone to scraped knuckles and possible broken connectors or scraped cables if you are not very careful. Picture of removed HV tank below:
Area
CT