GE Healthcare · CTLightSpeedVCT2601348052601347201002062260134806
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
Table will not move down; with tape or touch switch error See Long Description for switch locations. Table will not move down; tape switch was either contacted or not present The VCT table has a large number of touch sensors that can produce a variety of error codes.
Error message
Error: 260134805 IMS tape switch IN (rear) isn't present See Long Description for more Details Thu Nov 10 08:55:28 2005 Error: 260134805 Host: tgp Process: tgp_subsys File: gtcb_imc_isrfunc.cxx Line: 163 Thu Nov 10 08:55:28 2005 Host : Table Sub-System Ermes # : 260134805 Exception Class : Secondary Severity : Pri/Soft File : gtcb_imc_isrfunc.cxx Line# : 163 Function : No System Function Reported Scan Type : None/Unknown Scan: 405/2/0 IMS tape switch IN (rear) isn't present. Position is 84 um. Error: 260134720 Elevation touch switch4 (tape sw on underside of IMS side cover) isn't present. See Long Description Fri Oct 14 16:22:05 2005 Error: 260134720 Host: tgp Process: tgp_subsys File: gtcb_emc_hardware.cxx Line: 806 Fri Oct 14 16:22:05 2005 Host : Table Sub-System Ermes # : 260134720 Exception Class : Secondary Severity : Pri/Soft File : gtcb_emc_hardware.cxx Line# : 806 Function : No System Function Reported Scan Type : None/Unknown Scan: 367/1/0 Elevation touch switch4 (tape sw on underside of IMS side cover) isn't present. Position is 1002062 um from the table base. Error: 260134806 IMS tape switch OUT (front) isn’t present. See Long Description Thu Nov 10 08:51:41 2005 Error
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 5116570-3 GTTS BOARD ASSY Positioning GT (replaces 5116570) 5116570-3 GTTS BOARD ASSY Positioning GT Part #: 5116570-3 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: GTTS BOARD ASSY Positioning GT FRU: Yes Status: Orderable Comments: added per to ECR2094306 by sso#305010201 in Jan 07, 2010 212004269 08/15/2011 11:25 GMT
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required