Pular para o conteúdo principal
GE Healthcare · CTLightSpeed26011820924097

LightSpeed 5.X (60kW)

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 260118209 (60-0311) ILP current not OK See LD for error explanation Sun Nov 30 10:50:13 2008 Error: 260118209 Host: orp Process: orp_subsys File: gen_exec.c Line: 2446 Sun Nov 30 10:50:13 2008 Host : Jedi Ermes # : 260118209 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 2446 Function : No System Function Reported Scan Type : Helical Scan: 24097/2/1 (60-0311)ILP current not OK. The current in the parallel resonant circuit of the inverter did not rise at the beginning of the exposure. Z6,Z7= 0h,0h Message/explanation: ILP current not OK. The current in the parallel resonant circuit of the inverter did not rise at the beginning of the exposure.

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2246223-2 HV TANK HEMIT JEDI 60DC Used on 5.X 60KW and BrightSpeed 4/8/16 slice HPower Scanner Elite-16, Edge-8, Excel-4 See Long Description for details 2246223-2 HV TANK HEMIT JEDI 60DC Used on 5.X 60KW and BrightSpeed 4/8/16 slice HPower Scanner Elite-16, Edge-8, Excel-4 Part #: 2246223-2 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: HV TANK HEMIT JEDI 60DC (SPARE) FRU: Yes Status: Orderable Comments: Not Available D3186T PERFORMIX ULTRA TUBE UNIT Used on LS2.X, 3.X, 4.X, 5.X HP60, Hispeed QXi, DLS with H2 Gantry and DST with H2 Gantry. See Long Description for more info and Picture D3186T PERFORMIX ULTRA TUBE UNIT Used on LS2.X, 3.X, 4.X, 5.X HP60, Hispeed QXi, DLS with H2 Gantry and DST with H2 Gantry. Part #: D3186T Part Type: GP TUBES Part Weight: 184.31 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: PERFORMIX ULTRA TUBE UNIT FRU: Yes Status: Orderable Comments: ** FE ASSIST ** The Shipping weight of this Tube is 336 lbs = 152 kilogram. For LightSpeed 2X/3X/4X and HP60. For CTi order D3132T and Lightspeed 1X order D3176T (212001261 5/3/07) *** SECOND OPINION NOTES *** Perform Rotor Control Diagnostics--If tube rotor fails to reach RUN mode in three ACCEL steps look for: High Drive output currents Low Drive output voltages and Low Drive output frequencies. This indicates a load problem with tube. With rotor running turn off Gantry 120V power at PDU. Listen for excessive noise and short deceleration time. A good rotor will coast for several minutes without power. Rotor issues (errors 183316, 183318, 183319, 260118111, 260118177) caused by tubes tend to be more prevalent on .5 second scans. If tube is installed on Jedi system, the Anode cable tends to break down (hard errors). Intermittent HV Spits -- Run KV Loop, 140KV, 30 sec, Anode, Cathode and both. Listen to tanks with stethoscope or screw driver, any noise indicates bad tank. Any spits? Remove HV cables from tanks fill with 2inch of oil. Spits = No, could be tube or cables. Spits = Yes, bad tank. D3186T PERFORMIX ULTRA TUBE UNIT Used on LS2.X, 3.X, 4.X, 5.X HP60, Hispeed QXi, DLS with H2 Gantry and DST with H2 Gantry. 2401826-2 Jedi 60 Shielded Power Unit / High Voltage Tank Assembly FRU Package for G-DAS systems Includes: 2377668-2 Jedi 60 Power Unit 2186730 Jedi 60 HV Tank Shielding improved for G-DAS systems See LD for details 2401826-2 JEDI 60DC SHIELDED POWER UNIT / HV TANK HLB ASSY With FRU PACKING FOR G-DAS SYSTEMS (Backward compatible with M-DAS systems) Part #: 2401826-2 Part Type: GP CONSUMABLES Part Weight: 2.43 LBR Repairable: No Returnable: No Hazardous Material: N Description: JEDI 60DC SHIELDED POWER UNIT / HV TANK HLB ASSY With FRU PACKING FRU: Yes Status: Orderable Comments: Revised per to MFG confirmation. This is part of Jedi Collector # 2371333-2 INCLUDES: 2377668-2 JEDI 60DC HLB POWER UNIT with INVERTER and PPC KV Control 2186730 JEDI 60 STD HV TANK
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues