GE Healthcare · CTLightSpeed178526245061134200017861210010
LightSpeed 2.X (Plus, QX/i) - Octane
Symptom
Scanner Hardware Fault during Scout Scan data acquisition. see long description for details. Scanner Hardware Fault during Scout Scan data acquisition The error only occurred when the tube was in the area of home position (11:00) but it could be plus or minus 20-30 degrees, not exactly home.
Error message
Error: 178526 Scan Seq Id: 0 Write to mailbox with uuid 230 failed with return status 2103 See LD Mon Oct 6 07:06:08 2008 Error: 178526 Host: ct02_sbc Process: fwmgr File: CF.c Line: 900 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Write to mailbox with uuid 230 failed with return status 2103 Error: 245061 Process: scanmgr SCNMGR_ACQUISITION_DONE event was not received See LD for complete error Wed Feb 28 17:50:15 2007 Error: 245061 Host: ct99 Process: scanmgr File: norm_and_gen_timeouts.cxx Line: 90 Function: System Control : Scan Control Scan Type: Helical Scan: 0/0/0 Scan Seq Id: 1342000 SCNMGR_ACQUISITION_DONE event was not received Error: 178612 A certified mail request to the firmware failed See LD Mon Oct 6 07:06:08 2008 Error: 178612 Host: ct02_sbc Process: fwmgr File: CF.c Line: 1088 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 A certified mail request to the firmware failed. FW Requst: 10010 FW UUID: 230
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 2.X (Plus, QX/i) - Octane Always S-DAS 2.4.2_H2.4M5 LightSpeed 2.X (Plus, QX/i) - Octane (Pegasus or Rhapsode IG) - Always S-DAS Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 2.4.2_H2.4M5 FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 FEATURES/OPTIONS STD/OPTION Cat # Digital Laser Camera I/F B7700PD Smart Helical (s/w only) Standard B7530TK Fast Recon (s/w only) Option B7700JF BMD (requires B7710MP or MN) Option B7710MK Smart Prep Option B7530SP Smart Score ? ? 3D Analysis (Requires D3D) Option B7540TD Navigator (s/w only) Option B7540N DentaScan (s/w only) Option B7540SD Direct 3D (Requires B7710SA) Option B7810SS HIS/RIS Option (h/w and s/w) Option B7500PN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Area
- CT
Full cause & step-by-step resolution