Pular para o conteúdo principal
GE Healthcare · MR222384622238492223832222381722221382223823

TPS intermittently fails to reset due to defective TPS CPU.

Symptom

Unable to Download or Reset the TPS MGD or ISE Unable to download TPS or ISE or MGD chassis

Error message

2223846 Download status packet from TPS is not received by NSP Tue Feb 9 11:47:04 1999 Error: 2223846 Host: MRS2 Process: NSP File: TardisReset.m Line: 1363 Download status packet from TPS is not received by NSP.

2223849 TPS EPROM power-up boot failed. Mon Mar 15 07:26:38 2004 Error: 2223849 Host: MMCLX_2 Process: NSP File: TardisReset.cc Line: 986 TPS EPROM power-up boot failed.

2223832 IPG Loader Daemon is Not Up Error: 2223832 Host: lx-mr Process: NSP File: TardisReset.cc Line: 1292 IPG loader daemon is not up. Mon Mar 15 07:26:38 2004 Error: 2223832 Host: MMCLX_2 Process: NSP File: TardisReset.cc Line: 1292 IPG loader daemon is not up.

2223817 TPS is Not Responding. Please Reset TPS Wed Apr 28 16:58:01 2004 Error: 2223817 Host: GEMS Process: NSP File: TardisReset.cc Line: 398 TPS is not responding. Please reset TPS.

2222138 TPS IPG BOARD POWER-UP ERROR: INTERPROCESSOR TEST 2222138 "TPS IPG BOARD POWER-UP ERROR: INTERPROCESSOR TEST -- CPU was not found in dual ported RAM space. Interprocessor Test will not be attempted. Check IACK jumpers or missing boards between CPU and IPG board."

2223846 Download status packet from IPG is not received by NSP. Mon Mar 15 07:26:40 2004 Er

Applies to

Product
Contour 3 & 4
Software
Profile Contour 7.67 Applies to Profile1 through 4 and Contour1 through 4. This is the last revision of software for this product line. Requires 7.66 LFC. 7.67 is a patch.
Component
2100968 TPS MVME 162 SINGLE BD CONTROLLER TPS MVME 162 SINGLE BD CONTROLLER MOTOROLA SMM162-020GEMS USED ON VMX
Area
MR

Full cause & step-by-step resolution