Pular para o conteúdo principal
GE Healthcare · MRSigna2223847222381722196282232902

TPS Reset Failure with Processor Exception Frame Error Due to Dual PPC Module.

Symptom

Unable to Download or Reset the TPS MGD or ISE Unable to download TPS or ISE or MGD chassis

Error message

2223847 TPS boot status packet is not received by NSP See Long Description Tue Feb 9 11:47:44 1999 Error: 2223847 Host: MRS2 Process: NSP File: TardisReset.m Line: 1177 TPS boot status packet is not received by NSP.

2223817 TPS is Not Responding. Please Reset TPS Wed Apr 28 16:58:01 2004 Error: 2223817 Host: GEMS Process: NSP File: TardisReset.cc Line: 398 TPS is not responding. Please reset TPS.

2219628 The BAM Check Packet failed to reach the TPS Tue Jun 15 08:58:55 2004 Error: 2219628 Host: IRIS Process: scn File: SrxIPScreen.m Line: 1956 The BAM Check Packet failed to reach the TPS.

2232902 Processor Exception Frame, Bus Error: vector 2, interrupt level 0 Mon Nov 3 13:11:30 1997 Error: 2232902 Host: oc001 Process: NSP File: TPS:excCHECKER Line: 0 Processor Exception Frame, Bus Error: vector 2, interrupt level 0 current pc 0x279bf2, sr 0x3004, task name MC860_CEID_2, fault address 0x400b0008, fault status 0x105 additional information in /usr/g/service/log/vxExLg-4c752740

Applies to

Models
Signa
Product
HF OpenSpeed HFO1 HFO2 and HFO3 Only.
Software
HFO3
Component
SecondOpinion Dual PPC Module 2222532-3 See comments in resolution concept for second opinion requirements before ordering this part See comments in resolution concept for second opinion requirements before ordering this part
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required