Pular para o conteúdo principal
GE Healthcare · CTLightSpeed2600074565020126000745750084

LightSpeed 3.X (Ultra, Plus, QX/i) - Octane

Symptom

"Scan hardware Stopped Scan" with beam tracking errors. "Scan hardware Stopped Scan" with beam tracking errors

Error message

Error: 260007456 Beam Tracking Control Loop Error Exceeded The WARNING Tolerance. See LD for more details Error: 260007456 Beam Tracking Control Loop Error Exceeded The WARNING Tolerance Mon Mar 20 16:36:22 2006 Error: 260007456 Host: DAS Process: DCB Firmware File: zaxis.c Line: 578 Function: DATA ACQUISITION : Scan Data Acquisition Scan: 50201/5/19 Type: Axial Exception Level: Pri/Soft Time: 16:36:22:380 Log Series: 83 Beam Tracking Control Loop Error Exceeded The WARNING Tolerance. Tolerance : 200 umd Control Error : 319 umd Beam Side : A Side

Error: 260007457 Beam Tracking Control Loop Error Exceeded The ABORT Tolerance See Long Description for complete error Wed Apr 27 18:54:40 2005 Error: 260007457 6 DAS DCB Firmware zaxis.c 564 Function: DATA ACQUISITION : Scan Data Acquisition Scan: 50084/6/0 Type: None/Unknown Exception Level: Pri/Soft Time: 18:54:40:490 Log Series: 936 Beam Tracking Control Loop Error Exceeded The ABORT Tolerance. Tolerance : 400 umd Control Error : 448 umd Beam Side : B Side

Applies to

Models
LightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2159578-3 collimator control board used on 1X/2X/3X/4X/5X See Long description for details Part #: 2159578-3 Part Type: GP CONSUMABLES Part Weight: 2.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: COLLIMATOR CONTROL BOARD FRU: Yes Status: Orderable Comments: Not Available
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues