GE Healthcare · CTLightSpeedVCT1827005898218945200109124
LightSpeed 3.X (Ultra, Plus, QX/i) - GRE
Symptom
Step and Shoot (SmartStep or SmartView) hand controller not functioning See LD for more Step and Shoot (SmartStep or SmartView) hand controller not functioning When customer attempts to use the hand controller for VCT step and Shoot, the system will not initialize into step and shoot mode.
Error message
Error: 182700 Interventional Hand Held Control (IHHC) failed to enable See the long description for the full error code Mon Sep 19 10:35:34 2005 Error: 182700 Suite: CT01 Host: ETC Hand Held Control Process: Diag Exec File hc_e xec.c 1.38 No Method Line: 595 Function: Diagnostics OPERATOR INPUT/OUTPUT : Tableside Operator Interface Test Name: Unique ID: 58982 Iteration: 1 Exception Level: Hard Time: 18945 Error Type: Firmware Interventional Hand Held Control (IHHC) failed to enable. Verify IHHC is connected to scanner. If IHHC is not connected to scanner, connect IHHC to scanner. If IHHC is connected to scanner, disconnect IHHC from scanner and reconnect IHHC to scanner. Error: 200109124 Received BAD SmartStep Start Status From Firmware See Long Discription for Details Mon Jun 1 13:17:42 2009 Error: 200109124 Hst: lsct Process: scanRx File: uifIpcMgrScanRx.c Line: 3034 Function: Data Acquisition : OC Processing Received BAD SmartStep Start Status From Firmware, status = -1, seq_num = 532. Please check connection for the Hand Held Control
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - GRE Always 4 or 8 slice MDAS. H2_P_M4 H2_P_M8 LightSpeed 3.X - GRE Method of visual identification Small rounded gantry similar in appearance to LS 2.X and 4.X. Always has an MDAS4 or MDAS8 Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da Always has an MDAS4 or MDAS8 List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2199947 CT SONATA HAND HELD CONTROLLER CT SONATA HAND HELD CONTROLLER Part number 2199947 CT SONATA HAND HELD CONTROLLER Part number 2199947
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.