Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT260121011163362601210132669426012102120500

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

Gantry tilt moves a small distance then stops (less than 2 deg) Gantry tilts forward a small distance then stops

Error message

Error: 260121011 Tilt uncommanded motion error. See long description for details Mon Oct 1 04:50:33 2007 Error: 260121011 Host: tgp Process: tgp_subsys File: tilt_monitor.c Line: 719 Mon Oct 1 04:50:32 2007 Host : Table Gantry Processor Ermes # : 260121011 Exception Class : Secondary Severity : Pri/Soft File : tilt_monitor.c Line# : 719 Function : No System Function Reported Scan Type : Scout Scan: 16336/1/1 Tilt uncommanded motion error. Start position: 132 mdeg. Current Tilt position: -122 mdeg. Tolerance = 250 mdeg.

Error: 260121013 Gantry velocity less than the minimum allowed velocity. See Long Description for more details Wed Mar 29 11:28:24 2006 Error: 260121013 Host : Table Gantry Processor Ermes # : 260121013 Exception Class : Secondary Severity : Pri/Soft File : tilt mgr.c Line# : 1171 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Gantry velocity less than the minimum allowed velocity. Gantry velocity:-2 mdeg/sec Tilt position:-26694 mdeg Minimum allowable velocity:300 mdeg/sec

Error: 260121021 Tilt Stopping Error. Tilt did not reach target angle See LD for complete error Fri Dec 21 16:47:02 2007 Error: 260121021 Host: tgp Process: tgp_subs

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues